LTS420E-P Specifications
Temperature Range < -195°C to 420°C (add LNP95 cooling pump to cool sample below ambient)
Sample area of 58x43mm
Quick-release gas valves for atmospheric control
Swing out lid for easy sample loading
Can be used with transmitted or reflected light
Mounts directly to microscope table or substage
Stage body size - 169 x 110 x 25.8 mm
100 ohm platinum sensor
Temperature stability & accuracy to <0.01°C
Highly conductive metal for improved heat transfer
Maximum heating / cooling rate of 30°C/min
Response time of <1 second at 5°C/min at 50°C
Objective lens minimum working distance 8.2mm
Condenser lens minimum working distance 12.5mm